M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 131
Preventive Maintenance:
8 Preventive Maintenance
This section contains procedures for cleaning the M2540A and its components.
During preventive maintenance, look for conditions that may cause problems, and loose knobs
or worn TGC controls.
CAUTION Use adequate ESD safeguards when performing service and maintenance procedures. See “ESD
Safeguards” on page 139.
NOTE Do not use commercial cleaners to clean any surfaces of the ultrasound system. These
cleaners may cause the colors to fade. Use only mild soap.
Cleaning the Air Filter
The air filter resides inside the service panel at the right front of the lower enclosure, in front of
the fan. If the filter becomes very dirty, it can impede airflow, which may cause overheating of the
power supply or the e-box.
To remove and clean the air filter:
1. Remove the service panel at the right front of the lower enclosure. See “Front Service Panel”
on page 141.
2. Slide the filter up and out of the panel’s back.
3. Wash the filter with a spray of water from the side that faced the fan.