Philips M2540 User Manual Page 109

  • Download
  • Add to my manuals
  • Print
  • Page
    / 220
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 108
M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 109
Performance Tests: Resident Self Tests
There are three tabs in the RST window:
The Info tab shows software, hardware, and system information. See “Information Types” on
page 109.
The Test tab invokes various test modes and displays test status and results. See “Tests” on
page 111
The Log tab displays errors and test parameter data in the event viewer and displays service
log entries in the service log. See “Viewing Error Logs” on page 119. There is also a service
log on the Log tab, for service personnel to record information about the system. See “Using
the Service Log” on page 125.
Information
Ty p e s
The RST application provides information on the hardware and the software of the ultrasound
system. You can select the Info tab and choose specific information types in the tree control.
Choosing a tree item and pressing the Select key opens a drop-down list that contains Save Info
and Print Info functions. Information can be saved to a text file or printed on a connected plain
paper printer.
System Information
The RST application provides system information that includes the system serial number, the
institution name, installed software and hardware options, and installed peripherals. Figure 6-7,
on page 108 illustrates this display.
The software information provided includes the revision number and the revision date and time
of the operating system, the main ultrasound application, and the common off-the-shelf (COTS)
software.
Page view 108
1 2 ... 104 105 106 107 108 109 110 111 112 113 114 ... 219 220

Comments to this Manuals

No comments