M2540 Ultrasound System
Field Service Manual
M2540-92000-01 A
November 2002
Copyright © 2002 by Philips Medical Systems All rights reserved
1 About This Manual
About This
Manual
This manual provides information for troubleshooting and servicing
the Philips M2540A ultrasound system. Brief overviews for each
section in this book are listed below:
Section 1, “General Information, presents a product overview and a
description of major features, and lists applicable safety standards.
Section 2, “Specifications, describes the physical and electrical
specifications of the M2540A.
Section 3, “Safety, discusses safety issues pertinent to the ultrasound
system, and describes ESD precautions to be taken when servicing the
system.
Section 4, “Theory of Operation, includes a technical overview of
system functions. Functional descriptions are given for each system
circuit board. Functional block diagrams show the scanner and the
scan converter sections. This section also includes explanations of the
theories behind the electrical safety tests in the Performance Tests
section.
Section 5, “Installation, explains how to unpack and assemble the
system, how to install peripheral devices, and how to configure the
system’s parameters.
Section 6, “Performance Tests, comprises all tests and diagnostic
procedures that apply to the M2540A, including electrical safety tests.
The chapter emphasizes patient, operator, and service personnel
safety, and system safety. A procedure for verifying complete system
operation is included.
Page view 0
1 2 3 4 5 6 ... 219 220

Summary of Contents

Page 1 - Field Service Manual

M2540 Ultrasound SystemField Service ManualM2540-92000-01 ANovember 2002Copyright © 2002 by Philips Medical Systems All rights reserved1 About This Ma

Page 2 - Audience

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 10General Information: Preset FunctionsPreset FunctionsThe ultrasound system includ

Page 3 - Comments

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 100Performance Tests: Electrical Safety TestsFigure 6-5 Transducer Leakage Current

Page 4 - Assistance

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 101Performance Tests: Electrical Safety Tests5. Fill a test container with enough s

Page 5

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 102Performance Tests: Electrical Safety TestsValues exceeding the limits may indica

Page 6 - 1 General Information

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 103Performance Tests: Electrical Safety TestsFigure 6-6 Transducer Isolation Leakag

Page 7 - Description

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 104Performance Tests: Electrical Safety Tests2. Plug the analyzer into an available

Page 8 - System Monitor

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 105Performance Tests: PC DiagnosticsPC DiagnosticsPC problems can often be diagnose

Page 9

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 106Performance Tests: PC Diagnostics6. Follow the onscreen instructions.NOTE Ignore

Page 10 - Functions

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 107Performance Tests: Resident Self TestsResident Self Tests The M2540A Resident Se

Page 11 - Equivalencies

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 108Performance Tests: Resident Self TestsUser InterfaceYou can control the RST soft

Page 12 - Specifications

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 109Performance Tests: Resident Self TestsThere are three tabs in the RST window:• T

Page 13 - 2Specifications

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 11General Information: Keyboard EquivalenciesKeyboard EquivalenciesSome of the keys

Page 14 - Dimensions

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 110Performance Tests: Resident Self TestsHardware InformationThe hardware informati

Page 15 - 147.5 cm (58 in)

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 111Performance Tests: Resident Self TestsTestsThe following figure shows the RST Te

Page 16

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 112Performance Tests: Resident Self Tests• Extended Test: This is a comprehensive t

Page 17 - 53.5 cm (21 in)

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 113Performance Tests: Resident Self TestsTest ResultsThe test results are shown in

Page 18 - 53.45 cm (21 in)

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 114Performance Tests: Resident Self Tests➤ Closing the RST ApplicationYou can close

Page 19 - Electrical

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 115Performance Tests: Resident Self Tests➤ Running an Extended TestThe extended tes

Page 20

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 116Performance Tests: Resident Self TestsFigure 6-10 RST Extended Tests5. If you wa

Page 21 - Connection

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 117Performance Tests: Resident Self TestsTest ModulesEach subsystem test consists o

Page 22

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 118Performance Tests: Resident Self TestsThe two-digit codes that precede the submo

Page 23 - Audio/Video

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 119Performance Tests: Resident Self Tests➤ Viewing Error LogsThe RST application lo

Page 24 - External Video

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 12General Information: Safety SpecificationsSafety SpecificationsThis section descr

Page 25 - Video Output

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 120Performance Tests: Resident Self Testsapplications and subsystems affected, and

Page 26

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 121Performance Tests: Resident Self TestsRST Test LogThis log includes parameters o

Page 27 - Exposure

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 122Performance Tests: Resident Self TestsRST Log CodesThe Application Test Log’s Ca

Page 28 - Standard

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 123Performance Tests: Resident Self TestsTest Result DetailsTo view detailed test r

Page 29 - Index Selection

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 124Performance Tests: Resident Self TestsSystem LogThe Event Viewer system log incl

Page 30 - Explosive

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 125Performance Tests: Resident Self TestsUsing the Service LogThe service log on th

Page 31 - Connections

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 126Performance Tests: Resident Self Tests7. Right-click Service Log and select Save

Page 32 - Moving the

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 127Adjustments: Monitor Adjustment7 AdjustmentsMonitor AdjustmentThe four buttons o

Page 33 - Safety: Moving the System

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 128Adjustments: Monitor AdjustmentNavigating the OSD MenusThe four buttons on the f

Page 34 - Transporting

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 129Adjustments: Monitor Adjustment6. Use the two right-hand monitor buttons to scro

Page 35 - Compatibility

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 13Specifications:2SpecificationsThis section lists the specifications of the M2540A

Page 36 - Restrictions

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 130Adjustments: Monitor Adjustment15. Press the far left monitor button.Changing th

Page 37 - Electrosur

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 131Preventive Maintenance:8 Preventive MaintenanceThis section contains procedures

Page 38 - 4 Theory of Operation

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 132Preventive Maintenance:CAUTION Do not use any solvent other than water to wash

Page 39

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 133Preventive Maintenance:CAUTION Use proper ESD precautions for the vacuum itself

Page 40 - TR Boards

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 134Troubleshooting: PC Won’t Boot9 TroubleshootingThis section contains tips and pr

Page 41

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 135Troubleshooting: PC Boots, but Monitor is Blank5. Reset the PC power selector sw

Page 42 - BPAP Board

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 136Troubleshooting: Printer ProblemsPrinter ProblemsHP940 StartupThe first time the

Page 43 - Physio Module

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 137Troubleshooting: Electromagnetic InterferenceElectromag-netic InterferenceElectr

Page 44 - Connectors

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 138Service Procedures: Overview10 Service ProceduresOverviewThis chapter includes a

Page 45 - I/O Panel

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 139Service Procedures: Repair Philosophy• “Cables” on page 160• “Cart” on page 163R

Page 46

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 14Specifications: Physical DimensionsPhysical DimensionsThe tables and diagrams tha

Page 47 - Diagrams

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 140Service Procedures: Service AccessService AccessThis section describes how to ga

Page 48

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 141Service Procedures: Record KeepingRear Service DoorsTwo doors in the rear of the

Page 49 - Functional Block

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 142Service Procedures: PC ServicePC ServiceThis section contains procedures for ser

Page 50 - Token Ring Cable

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 143Service Procedures: PC ServiceNOTE The message “Press any key to boot from CD” d

Page 51

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 144Service Procedures: PC Service3. With the ultrasound application running, press

Page 52 - Test Setup

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 145Service Procedures: PC Service6. Click Select All.7. Click Start. The backup pro

Page 53 - 7ASW030-1

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 146Service Procedures: PC Service➤ Restoring Customer PresetsTo restore a customer’

Page 54

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 147Service Procedures: PC ServiceNOTE Options can be restored only to the system th

Page 55 - 7ASW030-2

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 148Service Procedures: PC Service• Video converter board (part of the VCR option)•

Page 56 - 7ASW030-4

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 149Service Procedures: PC Servicec. Slide the bezel to the right until it can be re

Page 57 - 7ASW030-5

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 15Specifications: Physical DimensionsFigure 2-1 System Front Dimensions ...

Page 58

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 150Service Procedures: PC Service➤ Removing the PC from the SystemTo remove the PC,

Page 59 - 5 Installation

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 151Service Procedures: E-boxE-boxYou can perform most e-box service without removin

Page 60

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 152Service Procedures: E-box7. Install the replacement board.8. Reverse step 5 thro

Page 61

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 153Service Procedures: E-box6. Reverse step 5 through step 1 to install the motherb

Page 62 - Monitor Signal

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 154Service Procedures: Power Supply4. Remove the eight screws that secure the distr

Page 63 - Installation: System Startup

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 155Service Procedures: Power Supply6. Remove the two screws that secure the power s

Page 64 - Configuration

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 156Service Procedures: Transducer Connector ModulesTransducer Connector ModulesThe

Page 65 - Localization

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 157Service Procedures: Monitor2. Remove the service panel at the right front of the

Page 66 - Figure 5-4 Top Border Window

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 158Service Procedures: Monitor Twivel6. Perform all RST and system checks to verify

Page 67

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 159Service Procedures: System Control Panel➤ Replacing the System Control PanelTo r

Page 68

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 16Specifications: Physical DimensionsFigure 2-2 System Side Dimensions 41.71 cm(16.

Page 69

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 160Service Procedures: Monitor Support and Front PanelMonitor Support and Front Pan

Page 70

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 161Service Procedures: CablesSystem Control Panel and Monitor CablesCables that con

Page 71 - Installation

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 162Service Procedures: Cables6. Remove the four P-clips from the speaker supports.7

Page 72

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 163Service Procedures: Cart Cart This section contains procedures for servicing the

Page 73 - Network

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 164Service Procedures: Cart➤ Replacing the Top Panel of the Lower Enclosure You can

Page 74

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 165Service Procedures: Cart14. Reverse step 13 through step 1 to install the lower

Page 75 - 2. Click Service

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 166Service Procedures: Cart4. Remove the PC. See “Removing the PC from the System”

Page 76

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 167Service Procedures: Cart1. Complete “Preparation” on page 165.2. Remove the two

Page 77 - ➤ Adding Server Information

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 168Service Procedures: Cart3. Remove the two screws that secure the cable clamp to

Page 78

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 169Service Procedures: CartSlides➤ Replacing the SlidesEach of the three lift colum

Page 79

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 17Specifications: Physical DimensionsFigure 2-3 System Rear Dimensions Range:129

Page 80

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 170Service Procedures: CartCAUTION Make sure none of the cables snags in the lower

Page 81

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 171Cabling:11 CablingThe following table lists the cables in the M2540A. The cables

Page 82

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 172Cabling:Keyboard line audio cable M2540-69060 1Power cordsUK power cord M2540-60

Page 83 - Hardware

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 173Cabling:VCR option cablesVCR option video in/out cable M2540-69070 2VCR option a

Page 84 - 2. Click Keyboard

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 174Configuration: Disk Drive Settings12 ConfigurationDisk Drive SettingsThe jumper

Page 85 - VCR configuration

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 175Configuration: VCR Switch SettingsCD Drive JumpersFigure 12-2 shows typical CD-R

Page 86

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 176Configuration: Sony B/W Printer Dip Switch SettingsSony B/W Printer Dip Switch S

Page 87 - Configuring for printers

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 177Parts: Overview13 PartsOverviewThis section contains lists of the M2540A’s parts

Page 88 - Training

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 178Parts: Major AssembliesTo learn about transducer parts, please see the M2540A Tr

Page 89 - 6Performance Tests

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 179Parts: Transducer Connector ModulesTransducer Connector ModulesThe following tab

Page 90 - Resistance Test

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 18Specifications: Physical DimensionsFigure 2-4 System Top Dimensions 39.4 cm(1

Page 91

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 180Parts: Internal PCInternal PCThe following figure shows the replaceable componen

Page 92 - Current Test

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 181Parts: Internal PCThe following table lists the replaceable components of the in

Page 93

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 182Parts: Internal PCOptional physio module Table 13-1110 Physio module M2540-62300

Page 94 - (Source)

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 183Parts: e-box and Power Supplye-box and Power SupplyThe following figure shows th

Page 95

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 184Parts: e-box and Power SupplyThe following table lists the components of the pow

Page 96 - Isolation

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 185Parts: Lower AssemblyLower AssemblyThe following figure shows the components of

Page 97

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 186Parts: Lower AssemblyThe following table lists the components of the lower assem

Page 98 - Tr a n s du c er

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 187Parts: Lower Assembly17 Lower I/O panel M2540-00670 1LAN bulkhead connector 1252

Page 99

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 188Parts: Lower Assembly— VCR option serial cable M2540-69090 1 Table 13-9— Printer

Page 100

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 189Parts: Lower Assembly— Nameplate Performance M2540-83020 1— System Serial Overla

Page 101

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 19Specifications: Electrical SpecificationsElectrical SpecificationsMonitor Table

Page 102

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 190Parts: Lift Column AssemblyLift Column AssemblyThe following figure shows the co

Page 103

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 191Parts: Lift Column AssemblyThe following table lists the components of the lift

Page 104

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 192Parts: Lift Column Assembly— Friction ring M2540-80010 2— Belleville washer 3050

Page 105 - Diagnostics

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 193Parts: Upper AssemblyUpper AssemblyThe following figure shows the components of

Page 106 - Checkit Pro

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 194Parts: Upper AssemblyTable 13-7 lists the components of the upper assembly shown

Page 107 - Resident Self

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 195Parts: Upper Assembly— Pivot mount assembly 1— Pivot mount M2540-00230 115 Pivot

Page 108 - User Interface

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 196Parts: Top ShelfTo p S h e l fThe following figure shows the components of the

Page 109 - Ty p e s

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 197Parts: Top ShelfThe following table lists the parts shown in Figure 13-6. Number

Page 110 - Hardware Information

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 198Parts: VCR Option PartsVCR Option PartsThe following table lists the parts of th

Page 111 - Available Tests

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 199Parts: Physio Option PartsPhysio Option PartsThe following table lists the parts

Page 112 - Te s t i n g O p t i o n s

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 2About This Manual: About This ManualSection 7, “Adjustments,” contains instruction

Page 113 - Service Log

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 20Specifications: Transducer SpecificationsTransducer SpecificationsThe following t

Page 114

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 200Parts: Exchange Part NumbersExchange Part NumbersSome parts are available at red

Page 115

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 201Parts: Optional PeripheralsOptional Peripherals The following table lists the pa

Page 116

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 202Parts: Supplies and AccessoriesSupplies and AccessoriesThe following table lists

Page 117 - Test Modules

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 203Parts: Supplies and AccessoriesPaperSony UPP-110S standard paper 5 roll box SNY

Page 118

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 204Parts: Supplies and AccessoriesBiopsy kitsBiopsy kit C3540 Civco 645-057Biopsy k

Page 119 - Ultrasound Application Log

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 205Transducers:14 TransducersTable 14-1 lists the transducers you can use with the

Page 120

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 206Glossary:15 Glossary2D mode A mode of operation where the display represents ref

Page 121 - RST Test Log

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 207Glossary:Display All the information visible on the system's monitor, inclu

Page 122 - RST Log Codes

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 208Glossary:Mmode Method of image processing in which the display is recorded as a

Page 123 - Test Result Details

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 209Glossary:Transducer The element of the ultrasound system that the sonographer pl

Page 124 - System Log

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 21Specifications: Connection and Communication SpecificationsConnection and Commu-n

Page 125 - Using the

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 1:About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 126

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 2:Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 127 - 7 Adjustments

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 3:Moving the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 128 - Ver ti ca l S ize

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 4:Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 129

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 5:➤ Powering On the System . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 130 - Boot Sequence

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 6:Chassis to Ground Resistance Test . . . . . . . . . . . . . . . . . . . . . . .

Page 131 - 8 Preventive Maintenance

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 7:➤ Launching the RST Application . . . . . . . . . . . . . . . . . . . . . . . . .

Page 132 - Preventive Maintenance:

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 8:Printer Installation Problems . . . . . . . . . . . . . . . . . . . . . . . . .

Page 133

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 9:E-box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 134 - 9 Troubleshooting

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 10:➤ Preparing the System for Servicing the Lift Column . . . . . . . . . . . . .

Page 135

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 22Specifications: Connection and Communication Specifications1. Only available with

Page 136 - Problems

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 11:Top Shelf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 137 - Interference

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 23Specifications: Physio Port SpecificationsI/O Panel PortsThe following table list

Page 138 - 10 Service Procedures

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 24Specifications: Audio/Video SpecificationsAudioExternal Video Table 2-8 Audio Spe

Page 139

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 25Specifications: Audio/Video SpecificationsVideo Output Specifications Table 2-10

Page 140 - Service

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 26Safety:3SafetyThis section provides an overview of safety considerations for the

Page 141 - Side Panels

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 27Safety: Transmit Power (Acoustical)Tr a n s m it Power (Acoustical)Acoustic outp

Page 142 - PC Service

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 28Safety: AIUM/NEMA Output Display Standard• Reduce exposure time, independent of a

Page 143

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 29Safety: AIUM/NEMA Output Display StandardNOTE The power index displayed on the sc

Page 144 - 4. Click the Floppy Disk tab

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 3About This Manual: About This ManualFormatThis manual is in Portable Document Form

Page 145

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 30Safety: Explosive HazardsWhen TIC is Displayed With Normal SettingIf none of the

Page 146

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 31Safety: Electrical Warnings• When using the imaging system in the operating room,

Page 147 - PC Components

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 32Safety: Glutaraldehyde ExposureGlutaral-dehyde ExposureThe United States Occupati

Page 148

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 33Safety: Moving the SystemWhen Moving the SystemThis system is equipped with a fro

Page 149

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 34Safety: Moving the SystemWARNINGUse care when tilting the system for an incline.

Page 150

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 35Safety: Electromagnetic CompatibilityWARNINGNever ride on a loading lift with the

Page 151 - Service Procedures: E-box

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 36Safety: Restrictions for UseRestrictions for UseThe M2540A ultrasound system is s

Page 152

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 37Safety: Electrosurgical UnitsElectrosur-gical UnitsElectrosurgical units (ESUs) a

Page 153

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 38Theory of Operation: Overview4 Theory of OperationOverviewThis section provides a

Page 154 - Power Supply

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 39Theory of Operation: Internal PCInternal PCThe M2540A uses an internal PC to coor

Page 155

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 4About This Manual: About This ManualCustomer Assistance Various support locations

Page 156 - Connector

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 40Theory of Operation: TR BoardsVCR-Option Components in the PC The VCR option incl

Page 157 - Monitor

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 41Theory of Operation: TR Boardsplugs into the distribution board. The two boards f

Page 158 - Control Panel

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 42Theory of Operation: BPAP BoardBPAP BoardThe beam processor/acoustic processor (B

Page 159

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 43Theory of Operation: Demodulator BoardDemodulator BoardThe demodulator board perf

Page 160 - Front Panel

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 44Theory of Operation: System MotherboardPrimary FunctionsThe module’s primary func

Page 161 - Monitor Cables

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 45Theory of Operation: I/O PanelTr a n s du c er Connector Module TypesThe four tr

Page 162 - Service Procedures: Cables

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 46Theory of Operation: Power Supplypart. Rotary and sliding knobs, rotary encoders

Page 163 - Service Procedures: Cart

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 47Theory of Operation: Functional Block DiagramsFunctional Block DiagramsThis secti

Page 164

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 48Theory of Operation: Functional Block DiagramsFigure 4-1 System Block Diagram wit

Page 165 - Preparation

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 49Theory of Operation: Functional Block Diagramse-box and Power Supply Functional B

Page 166 - Gas Spring

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 5About This Manual: About This Manual© 2002 Philips North America Corporation All R

Page 167 - Mechlok and Cable

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 50Theory of Operation: Functional Block DiagramsToken Ring CableThe token ring cabl

Page 168

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 51Theory of Operation: Functional Block DiagramsPhysio Functional Block DiagramFigu

Page 169

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 52Theory of Operation: Transducer Safety Testing: Test Setup and TheoryTransducer S

Page 170

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 53Theory of Operation: Transducer Safety Testing: Test Setup and TheoryFigure 4-4 T

Page 171 - 11 Cabling

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 54Theory of Operation: Transducer Safety Testing: Test Setup and TheoryInsertion de

Page 172 - Cabling:

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 55Theory of Operation: Transducer Safety Testing: Test Setup and TheoryFigure 4-5 T

Page 173

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 56Theory of Operation: Transducer Safety Testing: Test Setup and TheoryFigure 4-6

Page 174 - 12 Configuration

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 57Theory of Operation: Transducer Safety Testing: Test Setup and TheoryImposing an

Page 175 - Settings

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 58Theory of Operation: Transducer Safety Testing: Test Setup and Theory• If the tra

Page 176 - Printer Dip

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 59Installation: Overview5 InstallationOverviewThis chapter includes procedures for

Page 177 - 13 Parts

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 6General Information: Product Overview1 General InformationThis chapter includes th

Page 178 - Assemblies

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 60Installation: Assemble the System• Peripherals should mount toward the rear of th

Page 179

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 61Installation: Assemble the System1. Loosen the peripheral strap, and slide the pr

Page 180 - Internal PC

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 62Installation: System Startup4. Hook the bracket in a slot of the cart top or the

Page 181 - Parts: Internal PC

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 63Installation: System Startup3. Press the two center monitor buttons.The OSD Main

Page 182

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 64Installation: System Configuration9. Press the far left monitor button.10. On the

Page 183

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 65Installation: System ConfigurationLocalization➤ Setting the Localization Paramete

Page 184 - Parts: e-box and Power Supply

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 66Installation: System ConfigurationFigure 5-4 Top Border Window 4. Enter the insti

Page 185 - Assembly

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 67Installation: System ConfigurationFigure 5-5 Date and Time Properties Window6. Cl

Page 186 - Parts: Lower Assembly

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 68Installation: System ConfigurationFigure 5-6 Regional and Languages Options Windo

Page 187

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 69Installation: System Configuration• The locale selected in Standards and formats

Page 188

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 7General Information: System DescriptionThe system control panel includes four numb

Page 189

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 70Installation: System Configuration16. Click the Advanced tab.Figure 5-8 Regional

Page 190

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 71Installation: System ConfigurationOptions InstallationThe M2540A leaves the facto

Page 191 - Parts: Lift Column Assembly

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 72Installation: System ConfigurationFigure 5-10 Options Installation Window 3. Sele

Page 192

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 73Installation: System ConfigurationDICOM Network ConfigurationComplete the followi

Page 193

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 74Installation: System Configuration3. Click the This System tab.4. Click Network S

Page 194 - Parts: Upper Assembly

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 75Installation: System ConfigurationFigure 5-12 Setup Window (Options)2. Click Serv

Page 195

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 76Installation: System ConfigurationFigure 5-13 Field Service Options Window3. Clic

Page 196 - To p S h e l f

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 77Installation: System Configuration➤ Adding Server Information1. On the DICOM Setu

Page 197 - Parts: Top Shelf

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 78Installation: System ConfigurationNOTES• The value in the Name: field need not be

Page 198 - Software

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 79Installation: System ConfigurationFigure 5-15 DICOM Setup Window (Servers & R

Page 199 - Physio Option

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 8General Information: System DescriptionThe optional Physio module installs in one

Page 200 - Exchange Part

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 80Installation: System ConfigurationFigure 5-16 DICOM Setup (This System) 2. Click

Page 201 - Peripherals

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 81Installation: System ConfigurationFigure 5-17 Disk Full Strategy Window 3. Set th

Page 202 - Supplies and

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 82Installation: System ConfigurationFigure 5-18 DICOM Setup Window (Diagnostics) 3.

Page 203

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 83Installation: System ConfigurationSystem Hardware ConfigurationThe following proc

Page 204

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 84Installation: System Configuration➤ Assigning Option Buttons1. On the Setup windo

Page 205 - 14 Transducers

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 85Installation: System ConfigurationFigure 5-21 Define Option Keys Window 3. Click

Page 206 - 15 Glossary

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 86Installation: System ConfigurationFigure 5-22 Setup Window (Peripherals). 2. Clic

Page 207 - Glossary:

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 87Installation: System ConfigurationThe driver is reinstalled with the correct conf

Page 208

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 88Installation: Customer TrainingCustomer TrainingThe customer needs some instructi

Page 209

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 89Performance Tests: Overview6Performance TestsOverviewThis chapter includes all te

Page 210

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 9General Information: Optional Peripherals (VCR, Printers, and Image Devices)Physio

Page 211

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 90Performance Tests: Electrical Safety Tests• The ECG Lead Leakage Current test che

Page 212

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 91Performance Tests: Electrical Safety TestsWARNINGSome users routinely perform a s

Page 213

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 92Performance Tests: Electrical Safety Tests4. Make the appropriate connections bet

Page 214

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 93Performance Tests: Electrical Safety TestsFigure 6-2 Ground Wire Leakage Current

Page 215

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 94Performance Tests: Electrical Safety Tests4. Take ground wire leakage current mea

Page 216

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 95Performance Tests: Electrical Safety TestsFigure 6-3 ECG Lead Leakage Current Tes

Page 217

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 96Performance Tests: Electrical Safety Tests5. Take ECG lead leakage current measur

Page 218

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 97Performance Tests: Electrical Safety TestsFigure 6-4 ECG Lead Isolation Leakage C

Page 219

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 98Performance Tests: Electrical Safety Tests4. Connect ECG leads to the ultrasound

Page 220

M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 99Performance Tests: Electrical Safety TestsThe following tools are required for tr

Comments to this Manuals

No comments