M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 62
Installation: System Startup
4. Hook the bracket in a slot of the cart top or the peripheral tray so the bracket is as close as
possible to the printer.
5. Tighten the strap.
6. On the back of the printer, set dip switch number 5 to the UP position. All the other
switches should be down.
7. Switch the printer power switch on.
8. Save the following material from the printer’s shipping box:
Thermal paper
Cleaning sheet
Printer manual
Any other documentation
System
Startup
Complete the procedures in this section to set up the system for the customer’s use.
Powering On the System
1. Connect the system power cord to a wall outlet.
2. Press the power switch on the front of the system’s PC.
Monitor Signal
Voltage
When shipped from the factory, the monitor signal input voltage is set to 1.0 V. This must be
reset to 0.7 V, using the following procedure:
Setting the Monitor Signal Input Voltage
1. Disconnect the power cord from the rear of the monitor.
2. Press the two right-hand monitor buttons, and hold them in while reconnecting the
power cord.