M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 113
Performance Tests: Resident Self Tests
Test Results
The test results are shown in the list view on the right-hand side of the RST window. The display
shows the name of each test, the components it tested, and the number of times the component
passed and failed the test.
To view details of the test results, click the Log tab, and select the information type from the
Event Viewer tree. For information on errors generated during tests, see “RST Test Log” on
page 121.
When the tests are finished, the following message box opens:
Service Log
You can enter notes in the service log section of the event viewer, to record service procedures
you perform or problems you encounter. See “Using the Service Log” on page 125.
Launching the RST Application
You can launch the RST application as the PC boots up or while the ultrasound application is
running. The procedure is the same in both cases.
To launch the RST application when the system is fully booted:
1. Launch the ultrasound application.
2. Press the key combination Ctrl+Alt+R.
To launch the RST application during system boot-up:
Press the key combination Ctrl+Alt+R while the ultrasound splash screen is on the display.
When it launches, the RST application closes the ultrasound application.