M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 82
Installation: System Configuration
Figure 5-18 DICOM Setup Window (Diagnostics)
3. Under DICOM log contents, select No logging (disables logging).
4. Under Service log, select Enable Service logging.
5. Under System Log, make sure the Logging level is set to Error Level Logging.
6. Click OK to close the DICOM Setup window.