Philips M2540 User Manual Page 85

  • Download
  • Add to my manuals
  • Print
  • Page
    / 220
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 84
M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 85
Installation: System Configuration
Figure 5-21 Define Option Keys Window
3. Click the drop-down list for each option key, and select the appropriate functions.
NOTE The drop-down lists include only those options that are part of the purchased software
features.
4. Click OK.
VCR configuration
The system is configured at the factory for NTSC format. To reconfigure the system for PAL
format, complete the following procedure:
Configuring the System for PAL Format
1. On the Setup window, click the Peripherals tab
Page view 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 219 220

Comments to this Manuals

No comments