M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 106
Performance Tests: PC Diagnostics
6. Follow the onscreen instructions.
NOTE Ignore the error messages about missing SCSI disk controllers.
For detailed information on using DFT, go to:
http://service.boulder.ibm.com/storage/hddtech/DFT32-V300USER-DOC.pdf
NOTE When DFT testing is complete, set the boot sequence so that the CD-ROM drive is
first. (See “Changing the Boot Sequence” on page 130.)
Checkit Pro
Test PC functions using the CheckIt Pro software on the Diagnostic floppy disk.
Testing PC Functions Using Checkit Pro
1. Change the VGA display height adjustment to show the whole screen. See “Adjusting VGA
Vertical Size” on page 128.
2. Change the boot sequence so the floppy disk is the first item. (See “Changing the Boot
Sequence” on page 130.)
3. Insert the Diagnostic disk in the floppy disk drive.
4. Restart the system
5. Run the CheckIt Pro application.
6. When the CheckIt Pro testing is complete, enter the BIOS again, and make the CD-ROM
drive the first item in the boot sequence. (See “Changing the Boot Sequence” on page 130.)