Philips PR3840/00 User Manual Page 42

  • Download
  • Add to my manuals
  • Print
  • Page
    / 346
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 41
Pulse output parameters
Frequency range 1-100 Hz
Pulse duration 60-350 µs for programs 1 to 20
Maximum current output 60mA at 500-1000 Ohm
Maximum output voltage 60V
Current pulse shape
Biphase symmetrical (net current 0 DC)
Operating conditions
Temperature from +5°C to +40°C
Relative humidity from 15% to 93% (non-condensing)
Atmospheric pressure from 700hPa to 1060hPa
Storage and transport conditions
Temperature - electrodes from 0°C to +40°C
Temperature - device from -10°C to +50°C
Relative humidity less than 93% (non-condensing)
Atmospheric pressure from 700hPa to 1060hPa
Note: If you want to store the electrodes for more than
a month, keep them at temperatures between +5°C
and +27°C.
Note: Do not store the electrodes in the freezer or
refrigerator. Do not expose them to high temperatures,
nor immerse them in water, or leave them outside of
the plastic bag.
ENGLISH42