Printing the Error Log
To print the error log, complete the following steps:
1. Disconnect all transducers from the monitor and disconnect or switch off Telemetry.
2. Switch on the recorder.
3. Press and
Test at the same time:
C01 is shown in the US1/US display.
0 or 1 is shown in the Toco display.
4. Press again to select the Service Function Menu:
A01 (Print the Error Log) is shown in the US1/US display.
5. Press to print the error log. The display shows:
nnn nnn nnn
The following table lists the error log and permanent test functions, and their
corresponding menu numbers.
Figure 4-3 Example of an Error Log
Table 4-4 Service Function Menu Numbers and Error Log Functions
Menu No. Function
A01 Print the Error Log
A02 Clear the Error Log
A03 Start the Permanent Test
503 0001 91/05/12 17:33 91/05/12 17:33
510 0010 91/05/12 17:36 91/05/17 10:45
504 0001 91/05/12 17:33 91/05/12 17:33
506 0010 91/05/12 17:36 91/05/17 10:45
FIC
C1C4 0001 91/05/12 17:33 91/05/12 17:33
C146 000A 91/05/12 17:36 91/05/17 10:45