Transducers and Cables
Chapter 2 Technical Specifications 19
DECG Cable
(M1362A)
Cable weight: 22 grams
Cable Length: 666mm +/- 30mm
DECG Adaptor
Cable (M1362B)
Cable weight: 35 grams (approx.)
Cable Length: 77cm (approx.)
MECG Cable
(M1363A)
Cable weight: 16 grams
Cable Length: 606mm +/-30mm
M1364A With DECG
Cable M1362A
Patient Leakage Current: 120V at 60Hz, 10µA rms
Patient Auxiliary Current: < 0.1µA (dc)
Dielectric strength: 1500Vrms spark gap protected
Input Impedance: > 10M
CMRR: with patient cable, 51.5k/0.047µF imbalance at line frequency > 110dB
Noise: (referred to input with 25k) < 4µVp
Contact Potential Difference: ±500mV
Input Voltage Range: 20µVp to 3mVp
M1364A With
MECG Cable
M1363A
Patient Leakage Current: 120V at 60Hz, 10µA rms
Patient Auxiliary Current: < 0.1µA (dc)
Dielectric strength: 1500Vrms spark gap protected
Input Impedance: > 10M
CMRR: with patient cable, 51.5k/0.047µF imbalance at line frequency > 80dB
Noise: (referred to input with 25k) < 4µVp
Contact Potential Difference: ±500mV
Input Voltage Range: 80µVp to 4mVp
IUP Quartz
Transducer
(1290C option
J05)
Dynamic Range: -50 to 300mmHg.
Sensitivity: 5µV/V/mmHg.
Non-linearity: Whichever is greater: ±1% of the reading or ±1mmHg.
Volume Displacement: 0.2mm³/100mmHg.
Weight: 180g/6.3oz.
Cable Length: 3m/9ft 10in.
Length: 66mm/2.6in.