Chapter 10 System Checkout Procedures
141
Setting Value
Trigger Type (passive circuit
type)
Auto-Trak
Flow Trigger Sensitivity
(Active PAP or Active Flow
circuit type)
6.0 l/min
Leak Compensation (Active
Flow circuit type)
On
Flow Cycle Sensitivity
(Active PAP or Active Flow
circuit type)
20%
Rise Time 1
Ramp Length O
All other alarms O
B. Options Menu
Modify the settings in the Options menu to match those shown
below in Table 2.
Setting Value
Menu Access Full
Detailed View On
All other settings Discretionary
C. Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The
system will begin operating using the dened ventilation
settings.
Table 2 - Ventilator Settings
in the Options Menu