  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 4
The display
shows EExx
A system
error
occurred.
Retake the
measurement.
If the problem
persists, call
1-844-531-6861 for
assistance.
Data
transmission
or pairing
failed.
Bluetooth is
o.
Turn on Bluetooth
on your mobile
device.
Data
transmission
or pairing
failed.
The Philips
HealthSuite
health app
is o.
Press the icon on
your mobile device
to activate the
health app.
The blood
pressure
monitor and
mobile device
are more than
16 feet/ 5
meters away
from each
other.
Place your mobile
device closer to
the blood pressure
monitor.
You selected
the wrong
prole on
the blood
pressure
monitor.
Select the correct
user prole on the
blood pressure
monitor before
your measurement.
Otherwise the
data cannot be
transmitted to your
app. Repeat the
measurement with
the correct prole
selected
Specications
Product name
Philips Wrist blood
pressure monitor with
Bluetooth®
Power supply 3.7V 420mAH built-in
rechargeable li-polymer
battery
Display Display with white
backlight
Visible area = 1 13/16" (L)
x 1 3/16" (W) / 46 mm x
30 mm
Measurement method Oscillometric method
Measurement range Rated cu pressure:
0mmHg - 300mmHg
Measurement pressure:
40mmHg - 230mmHg
heart rate: 40-199 beats
per minute
Problem Possible
cause
Solution
The result
is dierent
when I
perform
mea sure-
ments on
my right
wrist.
The blood
pressure
monitor is
suitable to be
used on both
wrists, but
the measure-
ment results
on the right
wrist and
left wrist will
dier.
For a meaningful
comparison, try
to measure under
similar conditions
and measure on
the same wrist
every time.
The blood
pressure
monitor
does not
work when
I press the
'on' button
The
rechargeable
battery is
empty.
Recharge the
battery (see
'Charging').
The lock
switch is
set to the
'locked'
postion.
Set the lock switch
to the 'unlocked'
position (Fig. 7).
The light of
the display
dims and
a battery
symbol+Lo
is showing
The battery is
low.
Charge the battery
(see 'Charging').
The display
shows Err
Com mu ni ca-
tion error.
Check if the app
is on and try data
transmission again.
The display
shows E3
The cu is
not properly
secured.
Refasten the cu,
wait for 3 minutes
and then measure
again.
The display
shows E10
or E11
The device
detected
motion, tal-
king or the
heart rate
is too weak
during the
measureme-
nt.
Wait for 3
minutes and then
measure again.
Do not
move during
measurement.
The display
shows E20
The device
does not
detect the
heart rate.
Make sure the
device is in
contact with the
skin. Loosen the
clothing on the arm
and measure again.
The display
shows E21
The
measurement
failed.
Wait for 3 minutes
and then measure
again.
Problem Possible
cause
Solution
Fluctuations
of blood
pressure
during the
day are
normal.
Blood pressure
uctuates from
minute to minute
and normally
shows a circadian
rhythm over a 24-
hour period, with
highest readings in
the afternoons and
lowest readings at
night. That is why,
for comparable
measurements,
the measurements
should be taken at
approx. the same
time of day.
You are using
medication.
The variations in
blood pressure
can be greater
if you are using
medication.
You
performed
multiple
mea sure-
ments
directly after
each other.
Wait at least 3
minutes between
measurements.
This allows your
blood circulation to
recover.
My blood
pressure
mea sure-
ment from
the hospital
is dierent
from the
mea sure-
ment at
home.
Multiple
variables
may aect
your blood
pressure such
as the we-
ather, emo-
tions and
exercise.
Pay attention when
you measure your
blood pressure at
home. Check for
instance:
If the cu is not too
tight or too loose.
If the cu is
properly attached
on the wrist.
If you feel anxious
or stressed, try to
relax. Take a deep
breath 2-3 times
before you start a
measurement.
Advice: Rest for
5 minutes before
you measure your
blood pressure.
IMPORTANT: This warranty does not cover damage
resulting from accident, misuse or abuse, lack of
reasonable care, the axing of any attachment
not provided with the product or loss of parts or
subjecting the product to any but the specied
voltage.*
NO RESPONSIBILITY IS ASSUMED FOR ANY
SPECIAL, INCIDENTAL OR CONSEQUENTIAL
DAMAGES.
In order to obtain warranty service, simply go to
www.philips.com/support or call toll-free 1-844-
531-6861. It is suggested that for your protection
you return shipments of product by insured mail,
insurance prepaid. Damage occurring during
shipment is not covered by this warranty.
NOTE: No other warranty, written or oral, is
authorized by Philips Electronics North America
Corporation. This warranty gives you specic
legal rights, and you may also have other rights
which vary from state to state. Some states do not
allow the exclusion or limitation of incidental or
consequential damages, so the above exclusion
and limitations may not apply to you.
* Read enclosed instructions carefully.
Manufactured for: Philips Consumer Lifestyle,
A division of Philips Electronics North America
Corporation, P.O. Box 10313, Stamford, CT 06904.
PHILIPS and Philips Shield are registered
trademarks of Koninklijke Philips N.V.
Troubleshooting
This chapter summarizes the most common
problems you could encounter with the device.
If you are unable to solve the problem with the
information below, visit www.philips.com/support
for a list of frequently asked questions or call
1-844-531-6861 for assistance.
Troubleshooting
Problem Possible
cause
Solution
My blood
pressure
uctuates
throughout
the day.
Your meas-
urement
position, the
conditions
under which
you measure
or the time
of measure-
me nt, are
dierent
during each
measure-
me nt.
For a meaningful
comparison,
try to measure
under similar
conditions. For
example, take daily
measurements
at approximately
the same time, on
the same wrist, or
as directed by a
physician.